Short answer

GDPMD (Good Distribution Practice for Medical Devices) is the quality management system that Malaysia’s Medical Device Authority (MDA) requires from authorised representatives, importers and distributors; manufacturers need ISO 13485 instead. A conformity assessment body (CAB) registered with MDA under section 10 of Act 737 audits the company and issues the GDPMD certificate, which is submitted with the establishment licence application in MeDC@St. Since 1 July 2024 each role needs its own licence: RM 4,000 for an authorised representative, RM 2,000 for an importer or distributor, plus a RM 250 application fee. A foreign manufacturer does not get GDPMD itself; it needs a Malaysian Authorised Representative whose licence and GDPMD scope cover its device category.

What GDPMD is

MDA’s guidance on establishment licensing defines GDPMD as the regulatory requirement for authorised representatives, importers and distributors to establish, implement and maintain a quality management system for their activities throughout the medical device supply chain (MDA/GD/0027, second edition, section 3.5). The GDPMD requirements themselves are published by MDA as MDA/RR No. 1 (first revision, November 2015).

MDA explains the reason in its FAQ: the people and companies responsible for procurement, transport, delivery, storage, device tracking, installation, commissioning, service, maintenance and calibration need to be managed, because the risks of these activities can be similar to those in manufacturing, and poor control can affect the safety and performance of devices at the point of use.

Who needs GDPMD, and who needs ISO 13485

Every establishment must hold a quality management system certificate from a CAB registered with MDA before it applies for an establishment licence (MDA/GD/0027, section 4.5). Which certificate depends on the role:

Who needs which certificate
Role in MalaysiaQMS certificateEstablishment licence
ManufacturerMS ISO 13485:2017 / ISO 13485:2016Yes (manufacturer licence)
Authorised Representative of a foreign manufacturerGDPMDYes (AR licence)
ImporterGDPMDYes (importer licence)
DistributorGDPMDYes (distributor licence)
Outsourced warehousing or transport providerMay obtain GDPMDNot required; the distributor controls it and declares it in its quality manual
Tendering agent, leasing company—Not required (MDA/GD/0027, Annex A)

A foreign manufacturer is not an establishment under Act 737, because an establishment must be a company registered in Malaysia (section 4.2). It therefore does not apply for a licence or for GDPMD; it authorises a licensed Authorised Representative to act for it. MDA’s FAQ also confirms that a company placing its own brand on contract-manufactured devices is a manufacturer and needs ISO 13485, not GDPMD.

From GDPMD certificate to establishment licence

MDA/GD/0027 (Table 1) sets out the order of steps before an application:

  1. Confirm that the product is a medical device under section 2 of Act 737.
  2. Determine the type of establishment: manufacturer, authorised representative, importer or distributor.
  3. Establish, implement and maintain the QMS: GDPMD for authorised representatives, importers and distributors.
  4. Appoint a CAB registered with MDA under section 10 of Act 737 to assess the QMS.
  5. The CAB issues an audit report and a certificate of conformity.
  6. Apply for the establishment licence in MeDC@St 2.0+, uploading the audit report and certificate, the letter of authorisation, the details of the person responsible and a signed attestation.

Three practical points from MDA’s FAQ and announcements:

  • Local CAB only. MDA accepts QMS certificates for establishment licensing only from CABs registered in Malaysia under section 10 of Act 737. An audit by your EU notified body does not replace the certificate of a Malaysian establishment. (For device registration, your EU MDR certificate still matters: it opens MDA’s verification route.)
  • Provisional certificates. An establishment can apply for its licence once the CAB has issued a provisional GDPMD certificate, and must update MDA when the full certificate is issued.
  • Electronic licences. MDA announced the full implementation of electronic establishment licences from 16 April 2026.

The person responsible for the establishment must come from top management and be domiciled in Malaysia: a Malaysian citizen, or a non-Malaysian with an employment pass or a residential address in Malaysia (section 4.3).

One licence per role: types and fees

There are four establishment licences: manufacturer, authorised representative, importer and distributor, and each activity needs its own licence (MDA/GD/0027, section 5). Under MDA’s Single License Policy, combined multi-role licences can no longer be renewed from 1 July 2024; a company that acts as authorised representative, importer and distributor holds three licences.

Establishment licence fees
LicenceNew licence (RM)Renewal (RM)
Manufacturer4,0002,000
Authorised Representative4,0002,000
Importer2,0001,000
Distributor2,0001,000

An application fee of RM 250 applies to new licences and RM 200 to renewals (Fifth Schedule of the Medical Device Regulations 2012, as listed in MDA/GD/0027, section 9). MDA’s FAQ gives the effect of the policy: an AR and distributor now pay RM 6,000 for new licences instead of RM 4,000, and an AR, importer and distributor RM 8,000. Renewal must be applied for at least 90 days before the licence expires. CAB fees for GDPMD certification are set by each CAB.

GDPMD scope and changes

A GDPMD certificate lists the medical device categories it covers, and the scope matters. MDA’s FAQ is explicit: a company whose GDPMD scope covers only, for example, dental devices cannot register a device from another category until its scope covers that category. When an establishment receives an additional letter of authorisation, it informs its CAB, which audits it and issues an updated certificate with the new device category where applicable. A change of premises also needs a CAB re-audit and an updated certificate (MDA/GD/0027, sections 7.2 and 7.5).

What this means for a foreign manufacturer

For an EU MDR certified software or device manufacturer, GDPMD is something to check, not something to obtain. Only manufacturers and authorised representatives can apply for device registration, and the AR carries post-market obligations with you, so its licence and certificate are part of your market access. Before you sign:

  • Ask for the AR’s current establishment licence for the authorised representative role, not only an importer or distributor licence.
  • Ask for its GDPMD certificate and audit report, the name of the CAB, and check that the CAB is registered with MDA.
  • Check that the GDPMD scope covers the device category of your product. For software, ask the AR and its CAB how the category and the GDPMD procedures cover software delivery and updates.
  • Prepare the letter of authorisation on your letterhead using the template in Annex B of MDA/GD/0027, with the list of devices.
  • Agree in writing how complaints, incidents and field safety corrective actions are reported, and what happens if the AR is replaced: an AR that is terminated without replacement still carries post-market obligations for devices it placed on the market (section 4.6).
  • Remember that the AR cannot be the brand owner, and that you may appoint more than one AR for the same device (Annex A, case 7).

The registration route itself, including the CAB verification route for EU MDR certified devices, is covered in our Malaysia MDA registration guide, and labelling in the Malaysia labelling guide.

TrustedTraceMed does not act as an Authorised Representative, importer or distributor in Malaysia. Our Malaysia and Singapore registration support covers the route, the dossier and coordination with the partner you appoint.

Sources